Over two days, a laboratory's PCR assay reports several unexpected weak positives (Ct 36-38) in patients with low pretest probability. In the latest run, the no-template control amplifies at Ct 37, while the positive and extraction controls perform as expected. What is the most likely cause and best response?
1Amplicon contamination; hold results, decontaminate, and retest✓ 정답
2Loss of efficiency in the positive control; recalibrate and keep the run
3A true community outbreak; report all positives and notify infection control
4Inhibitors in patient specimens; dilute the extracts and report the results
해설
Amplification in the no-template control, together with a cluster of late-Ct positives in unlikely patients, indicates contamination, most often amplicon carryover. Affected results should be held, work areas and equipment decontaminated, unidirectional pre- to post-amplification workflow reviewed, and specimens retested with fresh reagents. Inhibition causes false negatives, not positive no-template controls.