Analytical Principles & Instrumentation · 분석 원리와 기기
In an assay that follows Beer's law, a 100 mg/dL standard has an absorbance of 0.400. A patient serum was diluted by adding 1 part serum to 3 parts diluent, and the diluted sample has an absorbance of 0.300. What is the concentration in the original serum?
175 mg/dL
2133 mg/dL
3225 mg/dL
4300 mg/dL✓ 정답
해설
The diluted sample concentration is (0.300/0.400) x 100 = 75 mg/dL. One part serum plus three parts diluent is a 1:4 dilution, so the original concentration is 75 x 4 = 300 mg/dL. Multiplying by 3 confuses the parts of diluent with the dilution factor, and 133 mg/dL comes from inverting the absorbance ratio.