Analytical Principles & Instrumentation · 분석 원리와 기기
In an LC-MS/MS method for quantifying a drug, a stable isotope-labeled form of the analyte is added to every calibrator, control, and patient specimen. What is the main purpose of this step?
1To correct for extraction losses and ion suppression or enhancement✓ 정답
2To serve as the single calibrator for the measuring range
3To remove phospholipids from the specimen before injection
4To raise the analyte mass above the instrument's detection limit
해설
An isotope-labeled internal standard behaves almost identically to the analyte during extraction, chromatography, and ionization but is distinguished by mass, so quantitation uses the analyte-to-internal-standard ratio. This compensates for variable recovery and matrix effects such as ion suppression. It does not replace the calibration curve or clean up the specimen.