Analytical Principles & Instrumentation · 분석 원리와 기기
A healthy 40-year-old man has a persistently increased serum hCG on a two-site sandwich immunoassay. Imaging shows no tumor, urine hCG is negative, and serial dilutions of the serum give nonlinear results. What is the most likely cause?
1Heterophile antibodies bridging the capture and detection antibodies✓ 정답
2High-dose hook effect from an extremely high hCG concentration
3hCG secreted by an occult testicular germ cell tumor
4Biotin from supplements binding the streptavidin solid phase
해설
Heterophile or human anti-animal antibodies can link the capture and labeled antibodies without analyte, creating a false positive sandwich signal. These large IgG antibodies do not pass into urine, so urine hCG is negative, and the false signal does not dilute linearly. Biotin and the hook effect lower sandwich results rather than raising them.