Analytical Principles & Instrumentation · 분석 원리와 기기
A patient in diabetic ketoacidosis has a serum creatinine of 2.4 mg/dL by the Jaffe (alkaline picrate) method but 1.1 mg/dL by an enzymatic method on the same specimen. What is the most likely explanation?
1Hyperglycemia inhibits the enzymes of the enzymatic method
2Acetoacetate reacts with alkaline picrate and falsely raises creatinine✓ 정답
3Only the Jaffe method detects an acute fall in GFR
4Ketones cause negative bias in the enzymatic method
해설
Acetoacetate is a noncreatinine chromogen that reacts with alkaline picrate, causing a positive bias in the Jaffe method; enzymatic creatinine methods are much less affected. Both methods measure the same analyte, so a real change in GFR would not produce a method-specific difference.