When a broth microdilution panel is read, the growth control well (no antimicrobial) is clear, and all drug wells are also clear. What is the correct interpretation?
1The isolate should be reported as having MICs at the lowest dilution tested
2The isolate is susceptible to every drug on the panel
3The test is invalid and must be repeated, because the inoculum did not grow✓ 정답
4Only the drugs with breakpoints can be reported
해설
The growth control must show adequate growth to prove the inoculum was viable and the medium supported growth; without it, clear drug wells cannot be interpreted as inhibition. The panel is repeated, often after checking inoculum preparation.