For a regulated analyte with an 80% passing score, a laborat… | 마이메르시 MyMerci
이런 문제가 486개 더 있어요도전 ›
Quality Assurance, Method Evaluation & Proficiency Testing · 정도보증·방법 평가·숙련도시험

For a regulated analyte with an 80% passing score, a laboratory scores 60% in the first proficiency testing (PT) event of the year, 100% in the second, and 60% in the third. How is this performance classified under CLIA?

해설
CLIA defines unsuccessful PT performance as unsatisfactory results in two consecutive events or in two of three consecutive events, so failing the first and third events qualifies even with a passing event between them. Each failure also requires documented corrective action, and there is no averaging across events or reset after a passing event.
같은 주제 다음 문제A laboratory's monthly indicator shows a sharp increase in hemolyzed chemistry specimens i…이 문제가 수록된 문제집미국 임상병리사 MLS(ASCP/ASCPi) 문제은행·CAT98,000원
MLS(ASCP/ASCPi) 7영역 문제은행 486 문제 · 로그인 없이 바로 볼 수 있어요

마이메르시로 국가고시 완벽 대비

기출문제와 상세 해설을 무료로. 내 약점을 분석하고 진도를 관리하며 더 똑똑하게 공부하세요.

무료로 시작하기

학습 참고용입니다. 실제 임상은 최신 지침과 소속 기관 프로토콜을 따르세요.