A chemistry section uses LIS autoverification. QC for the run is acceptable. Which patient result should the rules hold for technologist review instead of releasing automatically?
1Potassium 4.6 mEq/L on a specimen whose hemolysis index exceeds the assay limit✓ 정답
2Sodium 139 mEq/L, no instrument flags, delta check within limits
3Glucose 104 mg/dL, no instrument flags, no previous result on file
4Creatinine 0.9 mg/dL, no instrument flags, delta check within limits
해설
Autoverification rules should hold results with interference indices beyond the assay's acceptable limit, because hemolysis falsely raises potassium even when the number looks normal. Results without instrument flags, within the AMR, and passing delta checks (or with no prior result to compare) can be released automatically when QC is acceptable.