A burn patient has an impedance platelet count of 210 x 10^9/L, but the smear estimate is about 60 x 10^9/L. The smear shows numerous schistocytes and microspherocytes, and the platelet histogram does not return to baseline at its upper threshold. No platelet clumps are seen. What best explains the discrepancy?
1EDTA-dependent platelet clumping
2Red cell fragments counted as platelets✓ 정답
3Giant platelets excluded from the platelet channel
4Cold agglutinins causing red cell doublets
해설
Small red cell fragments and microspherocytes fall within the platelet size window on impedance counters, falsely raising the platelet count and lifting the upper end of the histogram. An optical or immunologic platelet count, or a careful smear estimate, should be reported instead. Platelet clumping and giant platelets both cause falsely low, not high, impedance platelet counts.