A Westergren erythrocyte sedimentation rate is set up in the laboratory. Which technical error would most likely produce a falsely increased result?
1The tube is left tilted a few degrees from vertical during the run✓ 정답
2The test is set up 12 hours after collection on a sample kept at room temperature
3The EDTA tube was underfilled, so the anticoagulant is in excess
4A refrigerated sample is tested without returning it to room temperature
해설
Tilting the Westergren tube lets red cells slide down one side while plasma rises along the other, accelerating sedimentation and raising the ESR. Long delays at room temperature, excess EDTA, and testing cold specimens all slow sedimentation and give falsely low results.