A patient receiving intravenous lipid emulsion has a grossly lipemic EDTA sample. Results: RBC 3.70 x 10^12/L, hemoglobin 15.8 g/dL, hematocrit 33.0%, MCV 89 fL, MCH 42.7 pg, MCHC 47.9 g/dL. No agglutination is seen on the smear. Which approach best corrects the erroneous values?

해설
Lipemic turbidity adds to the absorbance read in the hemoglobin channel, so hemoglobin, MCH, and MCHC are falsely increased while the RBC count, MCV, and hematocrit are unaffected. Replacing the plasma with an equal volume of saline or diluent removes the turbidity, and the corrected hemoglobin is used to recalculate MCH and MCHC. Warming corrects cold agglutinins, not lipemia, and estimating hemoglobin from the hematocrit is not an acceptable substitute for a measured value.