학습 자료 둘러보기 →The external processor and internal receiver are connected via magnetic coupling. Exposure to strong magnetic fields, such as an MRI scanner, can cause demagnetization or rotation of the internal magnet, leading to device failure or misalignment.
Instruct the client to keep the external processor at least 6 inches (15 cm) away from sources of electromagnetic interference, including metal detectors, security screening arches, and large stereo speakers.
If an MRI is medically necessary, the implant must be explicitly identified as MRI-conditional, not MRI-safe. Specific protocols, including head bandaging to prevent magnet displacement, are required under strict radiologist supervision.