# A chemistry section uses LIS autoverification. QC for the run is acceptable. Which patient result should the rules hold for technologist review instead of releasing automatically?

> source: MyMerci (mymerci.kr)  
> url: https://mymerci.kr/pages/nclex_q.php?qn_id=729845  
> language: ko  
> subject: Laboratory Operations · 검사실 운영

## 문제

A chemistry section uses LIS autoverification. QC for the run is acceptable. Which patient result should the rules hold for technologist review instead of releasing automatically?

## 보기

1. Potassium 4.6 mEq/L on a specimen whose hemolysis index exceeds the assay limit **✔ 정답**
2. Sodium 139 mEq/L, no instrument flags, delta check within limits
3. Glucose 104 mg/dL, no instrument flags, no previous result on file
4. Creatinine 0.9 mg/dL, no instrument flags, delta check within limits

**정답: 1**

## 해설

Autoverification rules should hold results with interference indices beyond the assay's acceptable limit, because hemolysis falsely raises potassium even when the number looks normal. Results without instrument flags, within the AMR, and passing delta checks (or with no prior result to compare) can be released automatically when QC is acceptable.