# In an indirect antiglobulin test using polyethylene glycol (PEG) enhancement, why are the tubes not centrifuged and read after 37 C incubation?

> source: MyMerci (mymerci.kr)  
> url: https://mymerci.kr/pages/nclex_q.php?qn_id=729664  
> language: ko  
> subject: Blood Banking · 수혈의학

## 문제

In an indirect antiglobulin test using polyethylene glycol (PEG) enhancement, why are the tubes not centrifuged and read after 37 C incubation?

## 보기

1. PEG inactivates IgM antibodies before the antiglobulin phase
2. PEG lyses red cells when centrifuged at 37 C
3. PEG neutralizes complement needed for the antiglobulin phase
4. PEG causes nonspecific aggregation of red cells when centrifuged **✔ 정답**

**정답: 4**

## 해설

PEG removes water and pushes red cells together, so centrifuging before washing produces nonspecific aggregates that cannot be dispersed; tubes go directly to washing and the antiglobulin phase. PEG does not inactivate IgM or lyse cells, and it is not a complement neutralizer.